Troubleshooting
Troubleshooting Guides
6-6
Model 146
i
Instruction Manual
Thermo Fisher Scientific
Alarm - Pressure
High pressure indication
Check the pump for a tear in
the diaphragm, replace with
pump repair kit if necessary.
Refer to “Preventive
Maintenance” chapter.
Check that capillaries are
properly installed and
O-rings are in good shape.
Replace if necessary. Check
flow system for leaks. See
the “Servicing” chapter.
Alarm - Intensity
Defective detector
Replace detector.
Defective measurement interface
board
Replace measurement
interface board.
See “Cell Frequency High” and
“Cell Frequency Low or Zero” in
.
See “Cell Frequency High”
and “Cell Frequency Low or
Zero” in
Alarm - Oz Level 1
Check (Opt)
Does not reach the level 1 setting
See
for
troubleshooting information.
Oz Level 2 Check (Opt)
Does not reach the level 2 setting
See
for
troubleshooting information.
Oz Level 3 Check (Opt)
Does not reach the level 3 setting
See
for
troubleshooting information.
Oz Level 4 Check (Opt)
Does not reach the level 4 setting
See
for
troubleshooting information.
Alarm - O
3
Conc.
Concentration has exceeded range
limit
Check to insure range
corresponds with expected
value. If not select proper
range.
Concentration low
Check user-defined low set
point, set to zero.
Alarm - Motherboard
Status
Internal cables not connected
properly
Check that all internal cables
are connected properly.
Defective motherboard
Recycle AC power to
instrument. If still alarming,
replace board.
Alarm - Interface
Status
Internal cables not connected
properly
Check that all internal cables
are connected properly.
Defective measurement interface
board
Recycle AC power to
instrument. If still alarming,
replace board.
Alarm - I/O Exp Status
I/O expansion board was installed,
but is not currently installed.
Re-install board, if desired,
and recycle power.
Table 6-2.
Troubleshooting - Alarm Messages, continued
Alarm Message
Possible Cause
Action
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...