Preventive Maintenance
Pressure Regulator Capillary Replacement
Thermo Fisher Scientific
Model 146
i
Instruction Manual
5-7
3. Insert the new capillary and reconnect the fittings.
Figure 5-2.
Replacing the Ozonator Capillary
Pressure Regulator
Capillary
Replacement
Use the following procedure to replace the pressure regulator capillary
(
).
Equipment Required:
Pressure regulator capillary
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Unscrew the knurled fitting, separate the components, and remove the
capillary (
).
Capillary
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.