Operation
Service Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-99
The Set Defaults screen is used to reset the configuration values to factory
defaults.
●
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Gas Therm >
Set Defaults
.
The Factory Calibrate Oven Thermistor menu is used for calibrating the
permeation oven heater thermistor to either low point, high point, or set
defaults.
●
In the Main Menu, choose Service > Perm Oven Settings >
Factory
Cal Oven Therm
.
FACTORY CAL OVEN THERM:
>LOW POINT
HIGH POINT
SET DEFAULTS
OPER MODE DIAGS ALARM
The Calibrate Oven Thermistor Low Point screen is used to view and set
the permeation oven thermistor resistance. The low and high point screens
function the same way.
●
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Oven Therm >
Low
or
High Point
.
Set Defaults
Factory Calibrate Oven
Thermistor
Low and High Points
RESTORE DEFAULT SETTINGS:
FOR PERM OVEN GAS THERM
RESTORE
OPER MODE DIAGS ALARM
RESTORE DEFAULT SETTINGS:
FOR PERM OVEN GAS THERM
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
OPER MODE DIAGS ALARM
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...