Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-63
SELECT OUTPUT RANGE:
SELECTED OUTPUT: V ALL
CURRENTLY: 0-10V
SET TO: 0-5V ?
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Minimum Value screen is used to edit the zero (0) to full-scale (100)
value in percentages for the selected analog output channel. See
Table 3–3
for a list of choices. The minimum and maximum output value screens
function the same way. The example below shows the set minimum value
screen.
●
In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > Select Channel >
Set Minimum
or
Maximum Value
.
MINIMUM OUTPUT PERCENT:
SELECTED OUTPUT: V ALL
CURRENTLY: N/A %
SET TO: 0000.
5
% ?
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
Table 3–3.
Analog Output Zero to Full-Scale Table
Output
Zero % Value
Full-Scale 100% Value
Gas Conc
Zero (0)
10,000 ppm
O
3
Conc
Zero (0)
Range setting
Perm Conc
Zero (0)
10,000 ppm
Target Gflow
0.0
Gas flow controller setting
Actual Gflow
0.0
Gas flow controller setting
Target Zflow
0.0
Zero flow controller setting
Actual Zflow
0.0
Zero flow controller setting
Target Tflow
0.0
Total gas and zero flow
controller settings
Minimum and Maximum
Value
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...