Servicing
Photometer Lamp Replacement (Photometer option)
7-32
Model 146
i
Instruction Manual
Thermo Fisher Scientific
Figure 7-12.
Photometer Components
5. Carefully slide the lamp out of the photometer.
6. Carefully slide new lamp into place until it bottoms and tighten the
lamp clamp screw.
7. Plug the new photometer lamp cable into the LAMP connector on the
photometer board.
8. Re-install the instrument cover.
9. Reconnect power and turn instrument ON.
10. After lamp has stabilized (approximately 15 minutes), refer to the
“Photometer Lamp Voltage Adjustment” procedure that follows to
adjust the photometer lamp voltage.
Thermistor
Detector
Detector Retaining Screw
Photometer
Lamp Insulation
Lamp Retaining Screw
Lamp Heater
Lamp
Lamp Housing
Detector Housing