Permeation Oven Setup Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-29
The Permeation Rate screen is used to enter the permeation rate of the
permeation capsule used. Allowable values are 1.0 to 99999.9. If the values
selected are lower or higher than allowed, “TOO LOW” or TOO HIGH”
appears on the screen.
Note
Changing this value may affect allowable limits for permeation
concentration levels. It is advisable to check these settings.
▲
●
In the Main Menu, choose Permeation Oven Setup >
Permeation
Rate
.
PERMEATION RATE NG/NG/MIN:
CURRENTLY: 1000.0
SET TO: 01020.
0
?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Permeation Level 1-5 screens are used to set minimum and maximum
permeation levels. These levels can be selected from the Operation or Flow
Modes screens.
Note
When a gas is selected and mixed with the permeation oven output,
permeation levels cannot be selected.
▲
●
In the Main Menu, choose Permeation Oven Setup >
Level 1-5
.
PERMEATION CONC LEVEL 1
MIN: 0.100 PPM
MAX: 5.000 PPM
SET TO: 000001.20
0
PPM ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
Permeation Rate
Level 1-5
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...