Operation
Photometer Setup Menu
3-34
Model 146
i
Instruction Manual
Thermo Fisher Scientific
O3 BACKGROUND:
O3: 4.400
SET BKG TO: 0.0000 ?
©ª
INC/DEC
SAVE VALUE
OPER MODE DIAGS ALARM
The O
3
Coefficient screen allows the O
3
span coefficient to be manually
changed while sampling span gas of known concentration.
Note
The concentration value will show “ERROR” if the measured
concentration is not a valid span value (either higher than the selected
range, or 0 or lower).
▲
●
In the Main Menu, choose Photometer Setup > Calibration Factors >
O3 Coef
.
O3 COEFFICIENT:
O3: 600
SET COEF TO: 1.000
©ª
INC/DEC
SAVE VALUE
OPER MODE DIAGS ALARM
The Reset User Calibration Default screen allows the user to reset the
calibration configuration values to factory defaults.
●
In the Main Menu, choose Photometer Setup > Calibration Factors >
Reset User Cal Defaults
.
O
3
Coefficient
Reset User Calibration
Default
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...