Operation
Service Menu
3-102
Model 146
i
Instruction Manual
Thermo Fisher Scientific
PHOTOMETER SAMPLE GAS:
INTENSITY: 98425 Hz
NOISE: 1.4
OPER MODE DIAGS ALARM
The O
3
Response Coefficient screen is used to manually adjust the O
3
response coefficient for the photometer option. The value is used to adjust
how fast the ozonator drive is changed when under photometer control and
should usually be left at 1.000. The O
3
response coefficient is visible only if
the photometer option is installed and when the calibrator is in service
mode. For more information on the service mode, see “
Service Mode
”
earlier in the chapter.
●
In the Main Menu, choose Service >
O3 Response Coefficient
.
O3 RESPONSE COEFFICIENT:
CURRENTLY: 1.000
SET TO: 1.250?
©ª
CHANGE VALUE
SAVE VALUE
OPER MODE DIAGS ALARM
The Display Pixel Test is used to test the LCD display. The display pixel
test is visible only when the instrument is in service mode. For more
information on the service mode, see “
Service Mode
” earlier in the chapter.
●
In the Main Menu, choose Service >
Display Pixel Test
.
DISPLAY PIXEL TEST:
DURING TEST PRESS
OR
f
xx
TO EXIT,
TO TOGGLE
XXXXXx
BEGIN TEST
GO BACK TO MENU
OPER MODE DIAGS ALARM
O
3
Response
Coefficient
Display Pixel Test
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...