Servicing
Photometer Board Replacement (Photometer option)
Thermo Fisher Scientific
Model 146
i
Instruction Manual
7-35
Photometer Board
Replacement
(Photometer option)
Use the following procedure to replace the photometer board (
Equipment required:
Photometer board
Philips screwdriver
Use the following procedure to replace the photometer board.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the ozonator board. Refer to
3. Remove the cables from the board.
4. Remove the two screws from the top of the board.
5. Pop off the board from the bottom mounting studs and remove the
board.
6. To install the new photometer board, follow previous steps in reverse.
7. Reinstall the ozonator board.
Detector
Replacement
(Photometer option)
Use the following procedure to replace the photometer detector
(
).
Equipment Required:
Detector assembly
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...