Operation
Alarms Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-83
PHOT LAMP TEMPERATURE:
ACTUAL MIN: 50.0
o
C
SET MIN TO: 52.0
o
C ?
©ª
INC/DEC
SAVE VALUE
OPER MODE DIAGS ALARM
The Photometer Intensity screen displays the current lamp intensity and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 0 to 300,000 Hertz. If the intensity reading goes beyond either
the minimum or maximum limit, an alarm is activated. The word
“ALARM” appears in the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Phot Intensity
.
PHOTOMETER INTENSITY:
ACTUAL 98425 Hz
>MIN 45000 Hz
MAX 150000 Hz
OPER MODE DIAGS ALARM
The Minimum Photometer Intensity alarm limit screen is used to change
the minimum intensity alarm limit. The minimum and maximum
photometer intensity screens function the same way.
●
In the Main Menu, choose Alarms > Sample Flow >
Min
or
Max
.
PHOTOMETER INTENSITY:
ACTUAL MIN: 45000 Hz
SET MIN TO: 55000 Hz ?
©ª
INC/DEC
SAVE VALUE
OPER MODE DIAGS ALARM
Photometer Intensity
Min and Max Photometer
Intensity Limits
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...