Operation
Service Menu
3-92
Model 146
i
Instruction Manual
Thermo Fisher Scientific
CALIBRATE AMBIENT TEMP:
CURRENTLY: 32.3
o
C
SET TO: 030.
5
o
C ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Bench Temperature Calibration screen is used to view and set the
bench temperature sensor calibration. This temperature is used for
compensation of the photometer option’s ozonator measurement. The
temperature calibration is visible only if the photometer option is installed
and when the instrument is in service mode. For more information on the
service mode, see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Bench Temp Calibration
.
CALIBRATE BENCH TEMP:
CURRENTLY: 32.3
o
C
SET TO: 032.
5
o
C ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Analog Output Calibration menu is a selection of 6 voltage channels
and 6 current channels (if I/O expansion board option is installed) to
calibrate, and allows the user to select the calibration action zero or span.
The analog output calibration is visible only when the instrument is in
service mode. For more information on the service mode, see “
Service
Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
Bench Temperature
Calibration
Analog Output
Calibration
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...