Operation
Instrument Controls Menu
3-60
Model 146
i
Instruction Manual
Thermo Fisher Scientific
EXT SOLENOID SETTINGS:
>1 NOP GAS A
2 NOP GAS B
3 NOP GAS C
4 NOP GAS D
5 NOP GAS E
6 NOP GAS F
7 NOP OZONATOR
ª
OPER MODE DIAGS ALARM
The Logic State screen is used to change the I/O solenoid to either
normally open or normally closed.
●
Press
to toggle and set the logic state open or closed.
EXT SOLENOID SETUP:
>LOGIC STATE OPEN
INSTRUMENT STATE
OPER MODE DIAGS ALARM
The Instrument State menu allows the user to select the instrument state
that is tied to the selected solenoid output. A submenu lists signal types of
either alarm or non-alarm to choose from.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Ext Solenoid Settings > Select Solenoid >
Instrument State
.
CHOOSE SIGNAL TYPE:
>ALARMS
NON-ALARM
OPER MODE DIAGS ALARM
Logic State
Instrument State
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...