Operation
Service Menu
3-96
Model 146
i
Instruction Manual
Thermo Fisher Scientific
PERM OVEN SETTINGS:
>CAL GAS THERMISTOR
CAL OVEN THERMISTOR
PERM OVEN SELECTION
FACTORY CAL GAS THERM
FACTORY CAL OVEN THERM
OPER MODE DIAGS ALARM
The Calibrate Gas Thermistor menu is used for calibrating the permeation
oven gas thermistor using a water bath or known resistance.
●
In the Main Menu, choose Service > Perm Oven Settings >
Cal Gas
Thermistor
.
CALIBRATE GAS THERMISTOR:
>WATER BATH
KNOWN RESISTOR
OPER MODE DIAGS ALARM
The Calibrate Gas Thermistor Water Bath screen is used to view and set
the permeation oven gas temperature to a known temperature value.
●
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
Thermistor >
Water Bath
.
CAL GAS THERM (BATH):
CURRENTLY: 45.80
o
C
SET TO: 45.0
0
o
C
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
Calibrate Gas Thermistor
Water Bath
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...