Optional Equipment
Permeation Oven
9-6
Model 146
i
Instruction Manual
Thermo Fisher Scientific
2. From the Permeation Oven menu in the Service menu, select
Cal Oven
Thermistor
.
Figure 9-1.
Cal Oven Therm Resistor Screen
3. Enter the exact value of the attached resistor and press
to save the
offset.
4. Press
to return to the Permeation Oven menu.
5. Remove the resistor and re-attach the permeation oven to POJ3.
Setting Temperature with Water
Bath
Use the following procedure for setting the measure temperature with water
bath.
1. Remove the thermistor from the permeation tube oven, but leave the
thermistor connected to the measurement interface board. Insert the
thermistor into the water bath next to an NIST traceable thermometer
(if necessary, use an extension cable to reach).
2. Turn on the power to the water bath. Using an NIST traceable
thermometer with a resolution of
±
0.01
°
C, adjust the water bath
to 45
°
C.
SAMPLE 12:34 ALARM
CAL OVEN THERM (RESISTOR):
CURRENTLY: 4369 Ohms
SET TO: 04369 Ohms
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...