Calibration
Ozonator Control Check
Thermo Fisher Scientific
Model 146
i
Instruction Manual
4-3
Figure 4-1.
Photometer Calibration Flow Schematic
Ozonator Control
Check
Use the following procedure to perform an ozonator control check. Refer to
the ozonator control check flow schematic (
1. Connect the 146
i
OUTPUT port to the 49
i
PS INPUT port.
2. Set ozone concentration at different levels.
3. Compare the ozone concentration readings on the 146
i
and 49
i
PS.
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...