Servicing
Analog Output Calibration
Thermo Fisher Scientific
Model 146
i
Instruction Manual
7-21
Analog Output
Calibration
Use the following procedure to calibrate the analog outputs if a meter
reading in the “Analog Output Testing” procedure differed by more than
one percent or after replacing the optional I/O expansion board.
Equipment Required:
Multimeter
1. Connect a meter to the channel to be adjusted and set to voltage or
current as appropriate.
shows the analog output pins and
identifies the associated channels.
2. From the Main Menu, press
to scroll to Service, press
>
to scroll to
Analog Out Calibration,
and press
.
The Analog Output Cal screen appears.
Note
If Service is not displayed, refer to “
, then return to the beginning of this step.
Ground
16, 18, 19, 35, 37
Current Output Return
16, 18, 20, 22, 24
Table 7-5.
Analog Input Channels and Rear Panel Pin Connections
Input Channel
Pin
1
1
2
2
3
3
4
5
5
6
6
7
7
9
8
10
Ground
4, 8, 11
Table 7-4.
Analog Output Channels and Rear Panel Pin Connections, continued
Voltage
Channel
Pin
Current
Channel
Pin
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...