Servicing
Analog Output Testing
7-20
Model 146
i
Instruction Manual
Thermo Fisher Scientific
7. Check that the meter is displaying the full scale value. If the meter
reading differs by more than one percent, the analog outputs should be
adjusted. Refer to the “Analog Output Calibration” procedure that
follows.
8. Press
to reset the analog outputs to normal.
Figure 7-7.
Rear Panel Analog Input and Output Pins
Analog
Voltage
Outputs
Current Outputs
Analog
Voltage
Inputs
Table 7-4.
Analog Output Channels and Rear Panel Pin Connections
Voltage
Channel
Pin
Current
Channel
Pin
1
14
1
15
2
33
2
17
3
15
3
19
4
34
4
21
5
17
5
23
6
36
6
25