Operation
Program Menu
3-36
Model 146
i
Instruction Manual
Thermo Fisher Scientific
The Calibrate O
3
Coefficient screen is used to automatically adjust the O
3
coefficient and enter the span concentration. The display shows the current
O
3
concentration reading and the current O
3
range. The next line of the
display is where the O
3
calibration gas concentration is entered.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more accurate the calibration will be. To be most
accurate, use the 300-second averaging time.
●
In the Main Menu, choose Photometer Setup > Calibration >
Cal O3
Coef
.
CALIBRATE O3:
O3: 0.803
SPAN CONC: 0000.80
4
?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Program menu sets the Model 146i’s internal computer program to
conduct a series of calibration events at preset times and durations. Up to
32 calibration events can be programmed. Next Event is read only and
shows the status of the next event number and time, there is no setting
screen.
Each event can be assigned gas, span, ozonator and permeation levels,
photometer mode, and event set time. This function is only available when
the instrument is NOT in service mode. If the service mode is active when
the next cycle time occurs, the program will not be triggered. Turn service
mode OFF to run next program cycle time. For more information on the
service mode, see “
Service Mode
” later in this chapter.
Note
When events are running, none of the settings may be changed via
the front panel interface, except for the program cycle enable/disable,
which may be toggled to cancel the current program.
▲
●
In the Main Menu, choose
Program
.
Calibrate O
3
Coefficient
Program Menu
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...