Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-39
The Events 1-32 screens allow the user to turn events on or off and modify
the calibrator’s operational settings. These screens are similar to the
operation screen, but the numeric values cannot be changed, except
duration minutes. Duration minutes lets you set the duration of an event
and can be set by pressing
and using
enter a value in
minutes. Press
to save the value or
to leave the screen. User 1-
8 allows the instrument to control the user configured digital outputs as
part of the event. When set to NONE, it will leave that event alone (it stays
the same as it was before the event started). If set ON, it will turn user
outputs on. If set OFF, it will turn outputs to off.
●
In the Main Menu, choose > Program > Events >
Event 1-32
.
●
Press
or
to toggle and
to set following items.
EVENT 10:
>ENABLED: ? 32 MIN
GAS A ? 0.212 SCCM
SPAN 1 2314 PPM
PHOT OFF 0 PPM
PERM OFF 0 PPM
OZONE OFF 0 %
USER1 DOUT ON
ª
OPER MODE DIAGS ALARM
USER2 DOUT OFF
USER3 DOUT NONE
USER4 DOUT NONE
USER5 DOUT NONE
USER6 DOUT NONE
USER7 DOUT NONE
USER8 DOUT NONE
The Instrument Controls menu contains a number of items. The software
controls listed in this menu enable control of the listed instrument
functions.
●
In the Main Menu, choose
Instrument Controls
.
Events 1-32
Instrument Controls
Menu
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...