Operation
Diagnostics Menu
3-72
Model 146
i
Instruction Manual
Thermo Fisher Scientific
MOTHERBOARD VOLTAGES:
3.3 SUPPLY 3:3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.1 V
-3.3 SUPPLY -3.3 V
OPER MODE DIAGS ALARM
The Interface Board screen (read only) is used to display the current voltage
readings on the interface board.
●
In the Main Menu, choose Diagnostics > Voltages >
Interface Board
Voltages
.
INTERFACE BOARD VOLTAGES:
> 3.3 SUPPLY 3:3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
-15.0 SUPPLY -15.0 V
24.0 SUPPLY 24.0 V
PHOTOMTR 12.0 V
O3 LAMP 11.6 V
ª
OPER MODE DIAGS ALARM
The I/O Board screen (read only) is used to display the current voltage
readings on the I/O expansion board. This menu is only displayed if the
I/O expansion board option is installed.
●
In the Main Menu, choose Diagnostics > Voltages >
I/O Board
Voltages
.
I/O BOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-3.3 SUPPLY -3.3 V
OPER MODE DIAGS ALARM
Interface Board Voltages
I/O Board Voltages
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...