Operation
Service Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-97
The Calibrate Gas Thermistor Resistor screen is used to view and set the
permeation oven gas thermistor resistance to a known resistor value.
●
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
Thermistor >
Known Resistor
.
CAL GAS THERM (RESISTOR):
CURRENTLY: 3850 Ohms
SET TO: 0400
0
Ohms
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Calibrate Oven Thermistor screen is used to view and set the
permeation oven thermistor resistance to a known resistor value.
●
In the Main Menu, choose Service > Perm Oven Settings >
Cal Oven
Thermistor
.
CAL OVEN THERM (RESISTOR):
CURRENTLY: 3850 Ohms
SET TO: 0400
0
Ohms
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Permeation Oven Setpoint screen is used to select the permeation oven
state to not present, or the desired temperature choice of 30, 35, and 45 °C.
●
In the Main Menu, choose Service > Perm Oven Settings >
Perm Oven
Selection
.
Resistor
Calibrate Oven Thermistor
Permeation Oven Selection
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...