Operation
Flow Modes
3-14
Model 146
i
Instruction Manual
Thermo Fisher Scientific
Once selected the screen reports the new flow mode. The user must press
to accept and activate these settings. The message “SAVING” is
displayed for 3 seconds, then removed.
●
In the Main Menu, choose Flow Modes > Permeation Oven > Gas Mix
> Gas A-F >
Zero
or
Span 1
-
5.
GAS A SPAN SELECTION:
ZERO 0.000 PPM
SPAN 1 0.020 PPM
>SPAN 2 2.000 PPM
SPAN 3 0.000 PPM
SPAN 4 0.000 PPM
SPAN 5 0.000 PPM
MANUAL
OPER MODE DIAGS ALARM
FLOW MODE GAS A PERM MIX:
GAS CONC: 02.00 PPM
PERM CONC: 0.100 PPM
TFLOW: 1000 SCCM
SAVING
SAVE FLOW MODE
OPER MODE DIAGS ALARM
The Permeation Oven Gas Manual Flow screen shows the zero-air and gas
flows and is used to manually set them for the previously selected gas. If the
selected zero-air flow is lower or higher than the acceptable range, “TOO
LOW” or “TOO HIGH” appears on the screen.
●
In the Main Menu, choose Flow Modes > Permeation Oven > Gax Mix
> Gas A-F >
Manual.
MANUAL SELECTION:
>ZERO AIR 10000 SCCM
GAS 100.00 SCCM
OPER MODE DIAGS ALARM
Permeation Oven Manual
Flow
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...