C-Link Protocol Commands
Measurement Configuration
Thermo Fisher Scientific
Model 146
i
Instruction Manual
B-35
This command sets the ozonator target concentration under photometer
control, in ppb. This command is only available if the photometer option is
installed.
Send:
set o3 conc 500
Receive:
set o3 conc 500 ok
o3 setting
This command reports the ozonator target concentration under
photometer control, in ppb. The example below reports a setting of 100
ppb. This command is only available if the photometer option is installed.
Send:
o3
setting
Receive:
o3 setting 1.000E+02 ppb
range
This commands report photometer O
3
range. If the mode is incorrect, the
instrument responds with “can’t, wrong settings”. The example below
reports that the O
3
full-scale range is set to 5.0 ppm, according to
Table B–
8
. This command is only available if the photometer option is installed.
Send:
range
Receive:
range 6: 500E+03 ppb
set range
selection
This command selects the O
3
full-scale ranges, according to
Table B–8
.
The example below sets the O
3
full-scale range to 2.0 ppm.
Send:
set range 5
Receive:
set range 5 ok
Table B–8.
Standard Ranges
Selection
ppm
0 0.05
1 0.1
2 0.2
3 0.5
4 1.0
5 2.0
6 5.0
7 C1
8 C2
9 C3
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...