C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 146
i
Instruction Manual
B-11
alarm bench temp min
alarm bench temp max
These commands report the photometer bench temperature alarm
minimum and maximum value current settings. The example below reports
the bench temperature alarm minimum value is 15.0 °C. These commands
are only available if the photometer option is installed.
Send:
alarm bench temp min
Receive:
alarm bench temp min 15.0
set alarm bench temp min
value
set alarm bench temp max
value
These commands set the photometer bench temperature minimum and
maximum values to
value
, where
value
is a floating-point number
representing bias voltage alarm limits in degrees C. The example below sets
the bench temperature alarm maximum value to 35 °C. These commands
are only available if the photometer option is installed.
Send:
set alarm bench temp max 35
Receive:
set alarm bench temp max 35 ok
alarm conc o3 min
alarm conc o3 max
These commands report the photometer O
3
concentration alarm minimum
and maximum value current setting. The example below reports the O
3
concentration alarm minimum is .20 ppb. These commands are only
available if the photometer option is installed.
Send:
alarm conc o3 min
Receive:
alarm conc o3 min 20.0 ppm
set alarm conc o3 min
value
set alarm conc o3 max
value
These commands set the photometer O
3
concentration minimum and
maximum values to
value
, where
value
is a floating-point representation of
the concentration alarm limits in ppb. The example below sets the O
3
concentration alarm maximum value to 5. These commands are only
available if the photometer option is installed.
Send:
set alarm conc o3 max 5
Receive:
set alarm conc o3 max 5 ok
alarm intensity min
alarm intensity max
These commands report the photometer detector intensity alarm minimum
and maximum value current settings. The example below reports the
intensity alarm minimum value is 45000 Hz. These commands are only
available if the photometer option is installed.
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...