Operation
Service Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-91
CALIBRATE PRESSURE SPAN:
CURRENTLY: 753.0 mmHg
SET TO: 760.
0
mmHg ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Restore Default Calibration screen is used to reset the pressure
calibration configuration values to factory defaults.
●
In the Main Menu, choose Service > Pressure Calibration >
Set
Defaults
.
The Ambient Temperature Calibration screen is used to view and set the
ambient temperature sensor calibration. The ambient temperature
calibration is visible only when the instrument is in service mode. For more
information on the service mode, see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Ambient Temp Calibration
.
Restore Default
Calibration
Ambient
Temperature
Calibration
RESTORE USER DEFAULTS:
RESTORE
OPER MODE DIAGS ALARM
RESTORE USER DEFAULTS:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
OPER MODE DIAGS ALARM
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...