C-Link Protocol Commands
Diagnostics
B-14
Model 146
i
Instruction Manual
Thermo Fisher Scientific
Table B–3.
Alarm Trigger Values
Value
Alarm Trigger
0 Floor
1 Ceiling
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag
volt
mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3. Each voltage value is
separated by a space.
Send:
diag
volt
mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag
volt
iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2
event
event
params
event
= 1-32
This command reports the parameters programmed for the specified event
event
.
Send:
event 1 params
Receive:
event 1 params gas A span 1 ozon 3 perm 1 photo
internal
event status
This command reports the current event that is being executed 1-32. A 0
response indicates no event is currently running.
Send:
event
status
Receive:
event status 1
Diagnostics
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...