Operation
Alarms Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-79
AMBIENT TEMPERATURE:
ACTUAL MIN: 15.0
o
C
SET MIN TO: 15.1
o
C ?
©ª
INC/DEC
SAVE VALUE
OPER MODE DIAGS ALARM
The O
3
Lamp Temperature screen displays the current ozonator lamp
temperature and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 55 to 85 °C. If the O
3
lamp temperature reading
goes beyond either the minimum or maximum alarm limit, an alarm is
activated. The word “ALARM” appears in the Run screen and in the Main
Menu.
●
In the Main Menu, choose Alarms >
O3 Lamp Temp
.
O3 LAMP TEMPERATURE:
ACTUAL 68.8
o
C
>MIN 60.0
o
C
MAX 80.0
o
C
OPER MODE DIAGS ALARM
The Minimum O
3
Lamp Temperature alarm limit screen is used to change
the minimum O
3
lamp temperature alarm limit. The minimum and
maximum O
3
lamp temperature screens function the same way.
●
In the Main Menu, choose Alarms > O3 Lamp Temp >
Min
or
Max
.
O3 LAMP TEMPERATURE:
ACTUAL MIN: 60.0
o
C
SET MIN TO: 68.0
o
C ?
©ª
INC/DEC
SAVE VALUE
OPER MODE DIAGS ALARM
O
3
Lamp
Temperature
Min and Max O
3
Lamp
Temperature Limits
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...