Error and Warning Messages
45
EN
16
Error and Warning Messages
*These steps may only be performed by an authorized service technician.
Error message
Cause
Troubleshooting
Display turns off
1 acoustic signal, cyclical
• Communication error
• Incompatible software ver-
sions on the modules (soft-
ware updates incomplete)
• Check cable/plug/board and replace if necessary.*
• Reload software.*
Device error/Call service/Electronics
error
2x acoustic signals, cyclical
• Board, plug, cable defective
• Checksum error of the pro-
gram memory or an EEPROM
• Check plug, cables, boards and replace if neces-
sary.*
• Reload software.*
Device error/Call service/Electronics
error
3 acoustic signal, cyclical
• Pressure sensor exceeds max.
permissible limits
• Replace sensor.*
Device error/Call service/Motor error
5 acoustic signal, cyclical
• Motor blocked, defective, or
not connected
• Roller wheel defective
• Impulse transmitter defective
• Board defective
• Check roller wheel, must rotate.
• Check motor connection cable.*
• Replace motor.*
• Check plug, cables, boards and replace if neces-
sary.*
Device error/Call service/Keyboard
error
9 acoustic signal, cyclical
• Keyboard defective
• Keyboard not connected
• Main board defective
• Connect/replace keyboard.*
• Replace main board.*
Device error/Call service/ Calibra-
tion error
10 acoustic signal, cyclical
• Sensor not calibrated
• Calibrate sensor.*
Warning Messages
Cause
Troubleshooting
Insert tube set
3 acoustic signal, cyclical
• Attempted device start with-
out inserted or improperly
inserted tubing
• Insert tubing correctly.
• Realign pump head with reference centering block.
Critical overpressure
Actual pressure display flashes
1x acoustic signal, cyclical
• Safety threshold for pressure
exceeded
• Reduce pressure. The pump starts automatically
with the preset values once the pressure has been
reduced below the critical threshold value.
Remove tube set
• Device was switched on with
disposable tubing inserted,
which is the reason the sen-
sor could not be checked
• Remove tubing from pump head. Restart the device
without inserted disposable tubing.
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......