Operating the Device
EN
5.7
Using the "Wash" Function
To improve visibility within the joint, activate the
"Wash"
function.
• The
"Wash time"
can be set for 10 to 30 seconds at intervals of 5 (see 7 "User
• The
pressure increase
is adjustable from 10-100% but is also limited to a max.
value of 200 mm Hg (factory setting = 50%).
• The
flow increase
amounts to 0.5 l/min, limited to a max. of 3 l/min.
Starting the "Wash" function
1. While the device is running, press the
"Wash"
function field (Fig. 5-1 "Device
(12)
) on the touch panel.
2. The
"Wash"
function is activated for the duration set in the user menu.
3. The function display depicts
Wash
. This display supersedes the "normal" op-
erating displays depicted during the device function.
Stopping the "Wash" function
The pump automatically returns to the previously set operating values after the
specified
"Wash time"
has expired.
To end the
"Wash"
function before the set time expires, press the
"Wash"
func-
tion field of the touch panel again. This returns the pump immediately to the pre-
viously set operating values.
5.8
Using the "Turbo" Function
To improve visibility within the joint, activate the
"Turbo"
function. This is helpful
under difficult surgical conditions such as the so-called SAD (subacromial ar-
throscopic decompression).
CAUTION!
The "Turbo" function is not designed for arthroscopies in small joints.
• The
"Turbo time"
is 2 minutes and can be stopped any time.
• The
flow increase
amounts to 0.5 l/min and is depicted in the max. flow rate
display (Fig. 5-1 "Device Front",
(9)
).
• The
"Turbo"
function cannot be activated as long as the
"Wash"
function is
used.
• The
"Wash"
function cannot be activated as long as the
"Turbo"
function is
used.
Starting the "Turbo" function
1. While the device is running, press the
"Turbo"
function field (Fig. 5-1 "Device
(11)
) on the touch panel.
2. The function display of the device depicts
"Turbo"
. This display supersedes
the "normal" operating displays depicted during the device function.
Stopping the "Turbo" function
The
"Turbo"
function is automatically stopped after 2 minutes. While the device
is running, press the
"Turbo"
function field to stop the function before the
"Tur-
bo" time
expires. This returns the pump immediately to the previously set oper-
ating values.
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......