Function Test
27
EN
9
Function Test
WARNING!
Functional test
The functional test must be performed prior to each surgery.
WARNING!
Sterilize reusable instruments and tubing before surgery to prevent infections.
Check all the single-use/disposable items before removing them from the pack-
age to ensure that the packaging is intact and that the expiration date is still val-
id.
WARNING!
Original accessories
For your own safety and that of your patient, use only original accessories.
WARNING!
Device defect
Do not use this device if a defect is suspected or detected during the function
check. This also applies to obvious defects, especially defects and damage of the
power plug and power cable.
9.1
Preparing the Function Test
Check to make sure the fluid bags meet the listed specifications (see chapter 5.5
"Connecting the Fluid Bags") and are suspended properly.
9.2
Starting the Device
1.
Use the Stand
by/ON key to turn the device on.
2.
Check the charge status of the battery with the display indicator.
3.
Inserting cassette tube set into irrigation pump
When prompted by the function display, insert the cassette tube set see
chapter 5.4 "Inserting the Cassette Tube Set"). Insert the tube cassette into
the cassette slot of the device until you can hear it snap into place.
4.
Connect the tap spike (optional safe lock connection) with the fluid bag.
Check to make sure all tube connections are free of mechanical stresses and
are routed without snagging.
5.
Connect an instrument.
6.
Press the respective function field on the touch panel to select the corre-
sponding type of joint.
7.
Press the
START/STOP
key on the device after the display depicts
Start pump
8.
The rotation symbol is depicted on the function display. Wait until tube sys-
tem and instrument are completely filled with fluid. The tube filling phase
takes about 10 seconds.
9.
Stop the pump by pressing the
START/STOP
key again. The function display
switches to
Start pump
.
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......