Safety functions
23
EN
Ten cyclical warning signals can be heard. Turn the device off and back on after
approx. 10 seconds. If the error message is displayed again, the device may not
be used any longer. Make sure the device can no longer be operated until a qual-
ified service technician conducts the appropriate tests and repairs.
Tube detection error
If the tube detection is defective, the display shows the following alternating
messages:
Device error <-> Tube detection error
Roller wheel stops and 6 warning signals are emitted cyclically.
Turn the device off and back on after approx. 10 seconds have expired.
If the error message is displayed again, the device may not be used any longer.
Make sure the device can no longer be operated until a qualified service techni-
cian conducts the appropriate tests and repairs.
6.2
Safety and Warning Messages
During device operation, the function display depicts safety and warning mes-
sages. These messages refer to the handling and safety of the pump during use.
Incorrect placement of the tube cassette
WARNING!
As long as the tube cassette is not properly snapped into place, the function dis-
play depicts "Insert tube set."
The pump does not start and 3 warning signals are emitted when the
Start/Stop
key is pressed. Snap tube cassette properly into place and restart the device.
6.3
Exceeding Nominal Pressure
In case of arthroscopies, it is possible that temporary overpressures occur (up to
more than 300 mm Hg in extreme cases). The pump is therefore equipped with
safety control mechanisms.
Exceeding nominal pressure during opera-
tion
The pump reduces any possible overpressure within the joint.
Pressure > 30 mm Hg as nominal pressure
If the actual pressure exceeds the nominal pressure by 30 mm Hg, the device ac-
tivates after 5 seconds the roller wheel in reverse motion to reduce the overpres-
sure.
If the pressure is not reduced to below the nominal pressure plus 30 mm Hg
within 5 seconds, a warning sound is emitted.
Pressure >250 mm Hg
If the intra-articulate pressure exceeds 250 mm Hg for more than 5 seconds, the
motor is switched off to interrupt the irrigation process. The warning
Critical
Overpressure
is depicted on the function display.
CAUTION!
The message "Critical Overpressure" overwrites all other possible messages.
Overpressure warnings serve as information for the attending physician to re-
spond appropriately to the existing excess pressure. The intra-articular joint
pressure can be reduced by opening the outflow cannula. If possible, the attend-
ing physician should perform this task.
When the pressure drops below 250 mm Hg, the warning
Critical Overpressure
disappears and the pump resumes its normal functionality.
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......