22
Safety functions
EN
6
Safety functions
The electronic components continuously monitor the proper function of the de-
vice. Device malfunctions are indicated with audible warning signals, error mes-
sages, and/or the blocking of device functions. In case of error messages, the
display switches from the default display to the error display. A table listing a
summary of possible error and warning messages is provided in chapter 16 "Error
and Warning Messages"is illuminated for the duration.
6.1
Device Self-Test
After being switched on, the device performs a self-test of the sensors, the motor,
and electronic components. The following describes the messages for errors of
the individual modules or components.
Sensor error
The offset values are verified as part of the device check process that is executed
when the device is switched on. If an impermissible deviation or an error in the
pressure measurement electronic system is detected, the display alternates the
following indications:
Device error <-> Sensor error
3 warning signals can be heard.
Turn the device off and back on after approx. 10 seconds. If the error message is
displayed again, the device may not be used any longer. Make sure the device can
no longer be operated until a qualified service technician conducts the appropri-
ate tests and repairs.
Motor error
In case of a motor drive malfunction, the display shows the following alternating
messages:
Device error <-> Motor error
5 warning signals can be heard. Turn the device off and back on after approx. 10
seconds. If the error message is displayed again, the device may not be used any
longer. Make sure the device can no longer be operated until a qualified service
technician conducts the appropriate tests and repairs. The
Motor error
warning
can also occur during the operation of the device due to a jamming roller wheel.
Electronics error
In case of an electronic component malfunction, the display shows the following
alternating messages:
Device error <-> Electronics error
2 warning signals can be heard. Turn the device off and back on after approx. 10
seconds. If the error message is displayed again, the device may not be used any
longer. Make sure the device can no longer be operated until a qualified service
technician conducts the appropriate tests and repairs.
Keyboard error
In case of a malfunction of the buttons on the front foil, the display shows the fol-
lowing alternating messages:
Device error <-> Keyboard error
9 warning signals can be heard. Turn the device off and back on after approx. 10
seconds. If the error message is displayed again, the device may not be used any
longer. Make sure the device can no longer be operated until a qualified service
technician conducts the appropriate tests and repairs.
Calibration error
If a calibration error occurs, the display shows the following alternating messag-
es:
Device error <-> Calibration error
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......