36
Annual Inspection
EN
The layout of the test and its setup are depicted in Fig. 12-1.
1. Switch device on. Wait until the device check has finished.
2. Insert the cassette.
3. Suspend fluid bag at a height illustrated in Fig. 12-1 and connect bag with ir-
rigation tube.
4. Immerse the end with the Luer lock into the measuring cup.
5. Select a joint.
6. Set the nominal values:
•
Nominal pressure = 150 mm Hg
•
Max. flow = 1 l/min
7. Press the
START/STOP
key. Rotating arrow starts to turn.
8. Wait until the tube set is filled and a red dot appears in the rotating arrow.
9. Clamp off the instrument tube end in the measuring cup without stopping
the pump.
10. Empty the measuring cup.
11. Release the tube end. The pump is running.
12. Press the
START/STOP
key after 1 min. The measuring cup then contains ap-
prox. 1 l (±10%) of fluid.
The basic function test has been successfully completed once these values have
been reached.
Fig. 12-1 Setup of basic function test
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......