11.2
Cleaning the Device
1.
Use the
Stand
-
by/ON
key to turn the device off.
2.
Remove the power cable.
3.
Wipe the surface of the device with a soft cloth moistened with the surface
disinfectant (for example Meliseptol® rapid). The concentration of the used
disinfectant depends on the information provided by the manufacturer of
the disinfectant. Make sure moisture does not enter the device.
NOTE!
Do not sterilize the device.
11.3
Cleaning the Autoclaveable Remote Control
Recommended cleaners
Use only ph-neutral and mild cleaners to clean the remote control. Use a brush.
Follow the instructions of the manufacturer concerning concentration and resi-
dence times.
11.4
Disinfecting the Autoclaveable Remote Control
Use only ph-neutral or slightly alkaline disinfectants. Perform the disinfection ac-
cording to the information provided by the manufacturer of the disinfectant.
NOTE!
The autoclavable remote control must completely be rinsed with the disinfec-
tant to ensure disinfection takes place.
• Disinfect only cleaned wired remote control units.
• The wired remote control unit may be disinfected thermally and by machine.
• Recommended disinfecting procedure: Thermal disinfection.
11.5
Sterilizing the Autoclaveable Remote Control
The manufacturer has tested the remote control for 100 sterilizations*. Never ex-
ceed the number of sterilization cycles indicated by the manufacturer.
Only a cleaned, disinfected, and dry remote control may be sterilized with satu-
rated vapor. Please comply with the instructions of the operating manual includ-
ed with your autoclave.
Sterilization requirements for the autocla-
veable remote control
CAUTION!
These instructions do not guarantee the sterility of the product. Your hygiene
and infection control department or responsible organization continues to be re-
sponsible for validating sterilization and sterility.
The following sterilization methods have been validated by the manufacturer:
Steam Sterilization
Pre-Vac
Gravity
Sterilization Methods
Pre-vacuumed sterilizer
Gravity-displacement
sterilizer
Packaging
Wrapped or Unwrapped
Wrapped or Unwrapped
Cycle Time
5 min (273 °F/134 °C,
3 bar)
5 min (273 °F/134 °C,
3 bar)
Dry Time
10 min
10 min
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......