Guidelines and manufacturer's statement - electromagnetic compatibility
43
EN
a
The field strength of stationary transmitters such as base stations of wireless
phones and cell phones, ham radio operators, AM and FM radio and TV stations
can theoretically not always determined in advance. A study of the installation
site should be considered to determine the electromagnetic environment con-
cerning the stationary transmitter. If the measured field strength at the pro-
posed FLOSTEADY Arthroscopy Pump installation and FLOSTEADY Arthroscopy
Pump operation site exceeds the concordance levels listed above, the FLOSTEADY
Arthroscopy Pump should be monitored to document proper functionality and
operation as intended. If unusual performance characteristics are observed, ad-
ditional measures may be required such as changing orientation or the location
of the device FLOSTEADY Arthroscopy Pump.
b
The field strength should be less than 3 V/m for the frequency range of 150 kHz
to 80 MHz.
14.7
Recommended Safety Distances between Portable and Mobile
HF Telecommunications Devices and the FLOSTEADY Arthrosco-
py Pump
The safety distance d in meters [m] for transmitters with a max. rated output not
listed in the table above can be calculated by applying the corresponding formula
in the respective column. P is the max. rated output of the transmitter in watts
[W] according to the information provided by the manufacturer of the transmit-
ter.
Note 1: The higher frequency range applies for 80 and 800 MHz.
Note 2: These guidelines are probably not realizable in all cases. The distribution
and spread of electromagnetic quantities differs depending on the absorption
and reflection of buildings, objects, and people.
Recommended Safety Distances between Portable and Mobile HF Telecommu-
nications Devices and the FLOSTEADY Arthroscopy Pump
The FLOSTEADY Arthroscopy Pump is intended for use in an electromagnetic
environment where HF interferences are controlled. The user/operator of the
FLOSTEADY Arthroscopy Pump can contribute to lowering electromagnetic
emissions by complying with the minimum distance between portable and
mobile HF telecommunications devices (transmitters) and the FLOSTEADY
Arthroscopy Pump - depending on the output power of the communication
device listed below.
Rated output of
the transmitter
[W]
Safety distance based on the transmitting frequency [m]
150 kHz to
80 MHz
d = 1.2
P
80 MHz to
800 MHz
d = 1.2
P
800 MHz to
2.5 GHz
d = 2.3
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......