12.3
Pressure Measuring Test
The layout of the test and its setup are depicted in Fig. 12-2 "Setup of pressure
measuring test".
The pressure test checks the pressure chamber, the pressure sensors, and the
pressure measuring to ensure all elements are functioning properly. This test re-
quires the complete cassette tubing and a container filled with
water
. The height
of the water column (hydrostatic pressure) is used for measuring and then con-
verted to mm mercury column (mm Hg). The height of the water column above
the pressure chamber has to match the value of the actual pressure display after
conversion.
Conversion formula: p (cm H2O) x 0.74 = p (mm Hg)
1. Hold the instrument tube into a container filled with water.
2. Set the nominal values:
•
Nominal pressure = 50 mm Hg
•
Max. flow = 1 l/min
3. Start pump.
Warning:
The tubing behind the roller wheel must be filled completely with
fluid.
4. Stop the pump.
5. The actual pressure display (bar graph) shows approx. 0 mm Hg.
6. Hold the instrument tube at a level of approx. 67.5 cm above the cassette.
Make sure the tube segment between this point and the cassette is com-
pletely filled with fluid. Also make sure that none of the tube is higher than
the indicated value.
7. The height of the water column is used to calculate the generated hydrostatic
(water) pressure according to 67.5 cm H2O x 0.74 = 50 mm Hg.
8. The actual pressure display has to depict approx. 50 mm Hg (±10 mm Hg).
9. When varying the height of the tubing above the cartridge, the actual pres-
sure display changes accordingly.
The test of the pressure measurement is successfully completed when the values
of the actual pressure display correspond to those of the water column height.
Fig. 12-2 Setup of pressure measuring
test
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......