28
Function Test
EN
9.3
Performing the Function Test
Pump is connected to fluid bag and instrument (see chapter 9.2 "Starting the De-
vice"). The pump is turned on.
CAUTION!
The test has to be performed outside of a joint.
1. Select the following values:
•
Nominal pressure = 100 mm Hg
•
Max. flow = 0.8 l/min
2. Start pump by pressing the
Start/Stop
key (inflow valve at instrument is open
and fluid emerges from the tube). The tube is filled with liquid.
3. Close the inflow valve at instrument. The bar graph display shows approx.
100 mm Hg.
4. Stop the pump by pressing the
Start/Stop
key.
9.4
Checking the "Wash" Function
1. Open the inflow valve.
2.
Start
the "Wash" function by pressing the function field
Wash
on the device
or the corresponding key on the wired remote control. The pressure and flow
rate increase for the Wash function preset in the user menu are depicted in
the nominal pressure and max. flow displays.
3.
Stop
the "Wash" function by again pressing the function field
Wash
on the
device or the corresponding key on the wired remote control. The max. flow
and nominal pressure displays show the last set values.
9.5
Ending the Function Test
The function test has been successfully completed. The device is now ready for
use. Position the device as needed for surgery.
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......