Operating the Device
17
EN
In order to separate sterile from non-sterile areas, assign the following tasks to
the "non-sterile" or the "sterile" nurse or medical personnel.
Non-sterile tasks:
1. Open packaging of the cassette tube set.
2. Have the sterile personnel remove the tubes and the cassette from the pack-
age.
Sterile tasks:
1. Keep the Luer lock connection within the sterile area and hand the end with
the tap spikes (optional safe-lock connection with cap) and the tube cassette
to the "non-sterile" nurse or medical personnel.
2. Connect the Luer lock connector with the instrument (e.g. inflow cannula).
Non-sterile tasks:
1. Insert tube cassette into device.
Hold on to the pump when inserting the tube cassette. Use the grip edge
(4)
at the bottom of the device to ensure the tube cassette is inserted properly
with one hand. Hold the tube cassette
(2)
as depicted in Fig. 5-4 "Inserting
2. Slide your fingers underneath the bottom of the device up behind the grip
edge
(4)
and hold on to the pump at the same time while using thumb and
hand to insert the tube cassette until it snaps into place. The front of the tube
cassette
(3)
must be flush with the device front. To remove cassette, press 1x
on the triangle pictogram on the front of the cassette. Part of the tube cas-
sette pops out and the cassette can then be removed.
Fig. 5-4
Inserting the tube cassette
(1)
Cassette slot
(2)
Tube cassette
(3)
Front of tube cassette
(4)
Grip edge from underneath
(1)
(2)
(3)
(4)
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......