
B.5
Inputs and outputs
Table B-10
Channel 1 – output with HART
HMI menu ID Parameter
Description
Default value
[unit]
Value range
Access
level
2.4.1.1
Loop current mode
Select single point or multi-drop
HART communication.
Loop current
mode, 4..20 mA
● Off
● Loop current
mode, 4..20
mA
● Multi drop
mode, 4 mA
User
2.4.1.2
Active operation
Shows the application possibili‐
ties of your hardware. The kind
of operation depends on the wir‐
ing.
Active operation
not possible
● Active
operation not
possible
● Active
operation
possible
User
2.4.1.3
PV selection
Select process value for current
output.
Mass flow
For none function‐
al safety applica‐
tions (SIL):
● Mass flow
● Volume flow
● Standard
volume flow
● Density
● Reference
density
● Medium
temperature
● Fraction A
● Fraction B
● Fraction A %
● Fraction B %
● Control valve
For functional
safety applica‐
tions (SIL):
● Mass flow
● Volume flow
● Density
User
Default settings
B.5 Inputs and outputs
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
291