ID/
Icons
Diagnostic
Action
Comment
166
Standard volume flow
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Only hydrocarbon
167
Totalizer 1
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
168
Totalizer 2
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
169
Totalizer 3
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
170
Loop current
Loop current simulated. Disable 'Simulation' be‐
fore returning to normal operation
Only HART devices
172
Transmitter
FW invalid. A component does not have the ex‐
pected FW version. Start a product firmware
update to update the component version or re‐
place the component.
173
Sensor
FW invalid. A component does not have the ex‐
pected FW version. Start a product firmware
update to update the component version or re‐
place the component.
174
Display
FW invalid. A component does not have the ex‐
pected FW version. Start a product firmware
update to update the component version or re‐
place the component.
175
IO
FW invalid. A component does not have the ex‐
pected FW version. Start a product firmware
update to update the component version or re‐
place the component.
Diagnosing and troubleshooting
10.2 Fault codes and corrective actions
MASS 2100 & FC300 (FCT030)
202
Operating Instructions, 06/2017, A5E39789040-AA