ID/
Icons
Diagnostic
Action
Comment
128
Standard density
Value above alarm limit. Check process condi‐
tions. Align Upper alarm limit to normal process
conditions.
129
Standard density
Value above warning limit. Check process con‐
ditions. Align Upper warning limit to normal
process conditions
130
Standard density
Value below warning limit. Check process con‐
ditions. Align Lower warning limit to normal
process conditions.
131
Standard density
Value below alarm limit. Check process condi‐
tions. Align Lower alarm limit to normal process
conditions.
132
Standard volume flow
Value above alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter 'Upper alarm limit'
Only hydrocarbon
133
Standard volume flow
Value above warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter 'Upper warning limit'
Only hydrocarbon
134
Standard volume flow
Value below warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter 'Lower warning limit'
Only hydrocarbon
135
Standard volume flow
Value below alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter 'Lower alarm limit'
Only hydrocarbon
136
Totalizer 1
Value above alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter 'Upper alarm limit'
Diagnosing and troubleshooting
10.2 Fault codes and corrective actions
MASS 2100 & FC300 (FCT030)
198
Operating Instructions, 06/2017, A5E39789040-AA