WARNING
Hot, toxic or corrosive process media
Risk of injury during maintenance work.
When working on the process connection, hot, toxic or corrosive process media could be
released.
● As long as the device is under pressure, do not loosen process connections and do not
remove any parts that are pressurized.
● Before opening or removing the device ensure that process media cannot be released.
WARNING
Improper connection after maintenance
Risk of explosion in areas subject to explosion hazard.
● Connect the device correctly after maintenance.
● Close the device after maintenance work.
Refer to Cables and cable entries (Page 233).
9.4.1
Service information
Service information is information about the condition of the device used for diagnostics and
service purposes.
Service information parameters
The basic service information parameters are:
● Driver Current
● Pickup 1 Amplitude
● Pickup 2 Amplitude
● Sensor Frequency
● Process Media Temperature
● Zero Point Adjustment Auto/Manual
● Zero Point Offset Value
● Manual Zero Point
● Zero Point Standard Deviation
Service and maintenance
9.4 Maintenance and repair work
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
171