Signal input - channels 3 to 4
The Signal Input can be configured to either Dosing control, Totalizer reset, Remote zero adjust
or Force/Freeze output(s).
)&
2SHUDWLRQ0RGH
6LJQDO,QSXW
'HOD\7LPH
3RODULW\
3URFHHG
View
no.
Text
Options/Description
4
Operation Mode
Select the signal input functionality
5
Delay Time
Set the signal input delay time
6
Polarity
Set the signal input polarity
Gas Application wizard (menu item 1.5)
The Gas Application wizard will guide you through configuration of essential parameters for
measuring gas flow. As default Low Flow Cut-Off and Empty Tube Detection are disabled.
Actual volume flow is difficult to use with low pressure; hence we recommend to use Corrected
Volumeflow.
&RUU9ROXPHIORZ
)&
&RQILJXUH
(PSW\7XEH'HWHFWLRQ
(PSW\7XEH/LPLW
3URFHHG
'HIDXOW9DOXHV
View
no.
Text
Options/Description
2
Default Values
Select whether or not to use the default values
3-4
Configure
Configure empty tube detection and limit
5
Corr. Volumeflow
Configure settings for corrected volumeflow
Commissioning
6.3 Local commissioning
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
107