
ID/
Icons
Diagnostic
Action
Comments
62
Fluid temp. below limit
Increase the fluid temperature. If the failure con‐
tinues then contact your local Siemens repre‐
sentative
63
Fluid temp. above limit
Reduce the fluid temperature. If the failure con‐
tinues then contact your local Siemens repre‐
sentative
64
Frame temp. below limit
Increase fluid temperature and check that ambi‐
ent temperature is within specified limits. If the
failure continues then contact your local Sie‐
mens representative
65
Frame temp. above limit
Reduce fluid temperature and check that ambi‐
ent temperature is within specified limits. If the
failure continues then contact your local Sie‐
mens representative
66
Standard Deviation' above limit
Measurement continues with values from last
successful zero point adjustment. Improve con‐
ditions for automatic zero point adjustment and
repeat adjustment.
67
Zero Point Offset' above limit
Measurement continues with values from last
successful zero point adjustment. Improve con‐
ditions for automatic zero point adjustment and
repeat adjustment.
68
Zero point adjustment failed
Measurement continues with values from last
successful zero point adjustment. Improve con‐
ditions for automatic zero point adjustment and
repeat adjustment.
69
Empty Tube Limit' exceeded
Make sure that the sensor is filled with liquid and
that the liquid density is within the specified
'Empty Tube Limit'
70
Too little fluid in tube
Make sure that the sensor is filled with liquid
Diagnosing and troubleshooting
10.2 Fault codes and corrective actions
MASS 2100 & FC300 (FCT030)
192
Operating Instructions, 06/2017, A5E39789040-AA