2.4.4
Installation in hazardous areas
WARNING
Equipment used in hazardous locations
Equipment used in hazardous locations must be Ex-approved for the region of installation
and marked accordingly. It is required that the special conditions for safe use provided in the
manual and in the Ex certificate are followed!
Hazardous area approvals
The device is approved for use in hazardous area and has the approvals listed below. Special
conditions for safe installation and operation specified by each approval authority are included
in the relevant certificate.
ATEX:
FCT030 field mount transmitter (can be installed in Zone 1 for gas and Zone 21 for dust):
Certificate: SIRA 11ATEX1342X
II 2(1) GD
Ex db eb [ia Ga] IIC T6 Gb Ta = -40°C to +60°C
Ex tb [ia Da] IIIC T85°C Db
IECEx:
FCT030 field mount transmitter (can be installed in Zone 1 for gas and Zone 21 for dust):
Certificate: IECEx SIR 11.0150X
Ex db eb ia [ia Ga] IIC T6 Gb Ta = -40°C to +60°C.
Ex tb [ia Da] IIIC T85°C Db
EAC Ex
FCT030 field mount transmitter / -40°C ≤ Tamb ≤ +60°C
1Ex db eb ia [ia Ga] IIC T6 Gb
Ex tb [ia Da] IIIC T85°C Db
CSA:
Ex Canada:
FCT030 field mount transmitter (can be installed in Zone 1 for gas and Zone 21 for dust):
Certificate: cCSAus 2508628
Ex db eb ia [ia Ga] IIC T6 Gb
Ex tb [ia Da] IIIC T85°C Db
Safety notes
2.4 Use in hazardous areas
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
21