
ID/
Icons
Diagnostic
Action
Comment
195
CH2
Loop current in lower saturation. Check process
conditions or align limit to normal operation. Ad‐
just channel 2 parameter 'Lower range value'
If Operation mode is config‐
ured to Current output
196
CH2
Loop current in upper saturation. Check proc‐
ess conditions or align limit to normal operation.
Adjust channel 2 parameter 'Upper range value'
If Operation mode is config‐
ured to Current output
197
CH2
Cable break. Check channel 2 current output
cable connection.
If Operation mode is config‐
ured to Current output
198
CH2
Frequency too low. Check process conditions
or align limit to normal operation. Adjust chan‐
nel 2 parameter 'Lower range value'
If Operation mode is config‐
ured to Frequency output
199
CH2
Frequency too high. Check process conditions
or align limit to normal operation. Adjust chan‐
nel 2 parameter 'Upper range value'
If Operation mode is config‐
ured to Frequency output
200
CH2
Pulse overflow. Pulse output insufficient pulse
separation. Reduce pulses per amount, or re‐
duce pulse width, or increase amount
If Operation mode is config‐
ured to Pulse output
201
CH3
Loop current in lower saturation. Check process
conditions or align limit to normal operation. Ad‐
just channel 3 parameter 'Lower range value'
If Operation mode is config‐
ured to Current output
202
CH3
Loop current in upper saturation. Check proc‐
ess conditions or align limit to normal operation.
Adjust channel 3 parameter 'Upper range value'
If Operation mode is config‐
ured to Current output
203
CH3
Cable break. Check channel 3 current output
cable connection
If Operation mode is config‐
ured to Current output
Diagnosing and troubleshooting
10.2 Fault codes and corrective actions
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
205