9.4 Analysis processing section
9
-
9
FRA 51602
9.4 Analysis processing section
The SI prefixes (k, M, etc.) that are placed before units of measurement and the scales they represent are a
(10
-18
), f (10
-15
), p (10
-12
), n (10
-9
), µ (10
-6
), m (10
-3
), k (10
3
), M (10
6
), and G (10
9
).
Measurement accuracy (fixed range)
Measurement accuracy Relative ac Calibration accuracy
Relative accuracy
± (| Basic accuracy | + | Dynamic accuracy | + | Inter-range accuracy × N|),
where N is | number of measurement range steps between channels |.
Example: If the measurement ranges of each channels are 300 Vrms and 1
Vrms, N is 5 (300 V to 100 V to 30 V to 10 V to 3 V to 1 V).
Calibration accuracy
The accuracy of external equipment that is connected to the instrument, such
as a shunt resistor or probe, or the accuracy of the calibration standard
equipment.
Basic accuracy
Upper: gain (ratio); Lower: phase
Measurement range
(rms)
Frequency
≦
100 kHz
≦
200 kHz
≦
1 MHz
≦
2 MHz
600 V
±0.2 dB
±1.2°
-
-
-
300 V
±0.1 dB
±0.6°
-
-
100 V
±0.05 dB
±0.3°
-
-
30 V
:
30 mV
±0.01 dB
±0.06°
±0.025 dB
±0.15°
±0.1 dB
±0.6°
Conditions: At least 30 cycles of integration, fixed measurement range for both channels, and
both channels have the same range. The gain and phase error for when the signal
input is at the full scale of the measurement range for both channels.
*For the cells that contain only "–", the measurement range exceeds the maximum input voltage (see
"
9.2 Analysis input section
, ● Maximum input voltage"), so either the measurement is not possible
or there is no accuracy specification for it.
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...