4.4 Integration
4
-
18
FRA 51602
4.4 Integration
This instrument performs measurement processing as a discrete Fourier transform (DFT) in units of one
cycle of the measurement waveform. By nature, the DFT essentially removes noise and high-frequency
harmonics. However, if the signal component being measured is small relative to the noise, or if highly
accurate measurement is required, the accuracy of measurement can be improved by increasing the
integration period (the number of integrations).
White noise is reduced in rough proportion to the square root of the integration setting (number or duration
of integrations). Noise components other than the measurement frequency are also reduced by increasing
the integration setting. However, the time required for measurement increases in proportion to the number
of integrations.
■ Setting the integration mode
There are two integration modes: manual integration, in which the number of integrations is
predetermined, and automatic integration, in which the number of integrations is adjusted automatically
according to the measurement data.
Select
【
OSC
】-
[INTEGRATION MODE]
.
[FIX]
: Manual integration
[SHORT]
: Automatic integration (short time)
[MED]
: Automatic integration (medium time)
[LONG]
: Automatic integration (long time)
● Manual integration
The normal setting is
[FIX]
. The integration is performed according to the number of times or the
duration that is set in
[INTEGRATION]
.
● Automatic integration
The number of integrations is determined by monitoring the variance of the measurement results (DFT
results). The integration is performed repeatedly in the order SHORT < MED < LONG until the
variance becomes small. Clean data is thus obtained, but the measurement time is long.
When the signal-to-noise ratio decreases only within a particular frequency range within the sweep
range, it may be possible to obtain clean data in a short time as the result of automatic integration.
However, because the number of integration times is determined on the basis of a statistic (variance),
the integration may end while the signal-to-noise ratio is insufficient or the integration may be repeated
more times than is necessary, depending on the nature of the signal being measured. (The integration
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...