5.3 Input settings
5
-
7
FRA 51602
■ Over-level detection and measurement data
When an over-level detection occurs, the measurement data becomes NaN (Not a Number). The
frequency data for which over-level was detected is not plotted on the graph.
If an over-level is detected and measurement is stopped (
【
Input
】
–
[OVER DETECT]
–
[ACTION]
–
[SWEEP STOP]
is on), the data from the beginning of the sweep up to the time of the over-level
detection is retained. If measurement is not stopped (–
[SWEEP STOP]
is off), the measurement data for
the frequency for which over-level was detected becomes NaN. The measurement data that is stored in
the USB memory device also becomes "NaN".
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...