8.5 Performance test
8-12
FRA 51602
8.5.8 Analyzer dynamic range
This instrument
: Set the oscillator AC amplitude to 10 Vpk and the sweep frequency to from
10 Hz to 2 MHz.
Set the number of integrations to 300 cycles and the analysis mode to
CH2/CH1.
Measuring instrument
: (none)
Connections
: Connect the BNC short plug to the
CH1
connector. Connect the
OUTPUT OSC
to
CH2
.
Measurement method
: Perform upsweep measurements with the instrument, and read the
minimum gain values in the ranges from 10 Hz or more up to 1 MHz and
from 1 MHz or more to 2 MHz from the LCD screen.
After completing the measurements, change to the connections for
measuring the dynamic range of CH2 and set the analysis mode to
CH1/CH2. Perform a sweep measurement and read the minimum gain
values for each range from the LCD screen.
Connections for measuring the dynamic Connections for measuring the dynamic
range of CH1 range of CH2
Judgement
Connections
Analysis
mode
Frequency range
Minimum
measurement
value
Passing range
*1
CH1
dynamic range
CH2/CH1
10 Hz
≦
f < 1 MHz
_ _ _. _ dB
140.0 dB or more
1 MHz
≦
f
≦
2 MHz
_ _ _. _ dB
80.0 dB or more
CH2
dynamic range
CH1/CH2
10 Hz
≦
f < 1 MHz
_ _ _. _ dB
140.0 dB or more
1 MHz
≦
f
≦
2 MHz
_ _ _. _ dB
80.0 dB or more
*1: This value may not be satisfied due to the effects of electromagnetic noise in the environment or
other such factors. However, the test will be passed if most of the data that results from the sweep
measurement does satisfy the value.
OSC
CH1
CH2
FRA 51602
BNC-BNC
Coaxial cable
BNC short plug
OSC
CH1
CH2
FRA 51602
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...